Comparison of typical radiation doses and risks using an anthropomorphic ‘bone fracture’ phantom for commonly performed X‐ray projections in a 5‐year‐old

Doyle, Edel; Dimmock, Matthew R; Lee, Kam L; Thomas, Peter; Bassed, Richard B

Authors

Edel Doyle

Kam L Lee

Peter Thomas

Richard B Bassed



Abstract

Introduction: Diagnostic reference levels (DRLs) are typical dose levels for medical imaging examinations for groups of standard‐sized patients or standard phantoms for broadly defined types of equipment used as a tool to aid optimisation of protection for medical exposures. Currently, there are no paediatric DRLs for conventional radiography (i.e. general X‐rays) published in Australia. The aim of this study was to establish typical radiation doses and risks that are representative of those delivered for commonly performed X‐ray projections for a 5‐year‐old/20 kg child using a 5‐year‐old anthropomorphic ‘bone fracture’ phantom in three dedicated paediatric radiology departments in Victoria. Methods: A total of 20 projection images were acquired for a standard 5‐year‐old/20 kg phantom using digital radiography X‐ray equipment. The air kerma‐area product (KAP) measured at each centre by a KAP metre, which was calibrated to a national primary standard, was considered to represent the median value for that centre for each X‐ray projection. Organ doses and effective dose were estimated using PCXMC software, and risks of radiation‐induced cancer and radiation‐induced death were calculated based on the BEIR VII report. Results: The typical doses for the individual X‐ray projections ranged from 3 mGy•cm2 to 86 mGy•cm2, whilst the effective doses ranged from 0.00004 to 0.07 mSv. The radiation risks were ‘minimal’ to ‘negligible’. Conclusion: The estimation of typical radiation doses and associated risks for a 5‐year‐old/20 kg phantom study provides reference values for guidance and is a first step in assisting optimisation at other institutions until national DRLs, based on patient data from the clinical setting, are published.

Citation

Doyle, E., Dimmock, M. R., Lee, K. L., Thomas, P., & Bassed, R. B. (in press). Comparison of typical radiation doses and risks using an anthropomorphic ‘bone fracture’ phantom for commonly performed X‐ray projections in a 5‐year‐old. Journal of Medical Radiation Sciences, https://doi.org/10.1002/jmrs.717

Journal Article Type Article
Acceptance Date Aug 5, 2023
Online Publication Date Aug 21, 2023
Deposit Date Sep 7, 2023
Publicly Available Date Sep 7, 2023
Journal Journal of Medical Radiation Sciences
Publisher Wiley Open Access
Peer Reviewed Peer Reviewed
DOI https://doi.org/10.1002/jmrs.717
Keywords radiation effects, radiology, skeletal/bones, imaging phantoms, musculo skeletal, medical imaging, paediatric

Files

Comparison of typical radiation doses and risks using an anthropomorphic ‘bone fracture’ phantom for commonly performed X‐ray projections in a 5‐year‐old (112 Kb)
Archive

Licence
https://creativecommons.org/licenses/by-nc-nd/4.0/

Publisher Licence URL
https://creativecommons.org/licenses/by-nc-nd/4.0/

Copyright Statement
This is an open access article under the terms of the Creative Commons Attribution‐NonCommercial‐NoDerivs License, which permits use and distribution in any medium, provided the original work is properly cited, the use is non‐commercial and no modifications or adaptations are made.






You might also like



Downloadable Citations